Battery corpuls cpr (LiPo)
الأجهزة المتأثرة
REF: 09120 Serial numbers / Lot identification: 150000207 to 241500195 Software / Firmware: Each firmware version up to and including 1.4-005
الممثل المعتمد
Al Khateeb United Trading And Contracting Company
سبب المشكلة
Sporadically, the specified life span of the battery corpuls cpr (LiPo) REF 09120 is not reached. Cause for this is the cell voltage which falls below a certain voltage threshold due to deep discharging or longer storage periods if the battery is not charged regularly, as prescribed in the user manual. This promotes the progressive degradation of the electrode materials. This results in the battery going into error statuses and the life span is shortened.
الإجراء التصحيحي
- Refer to "Immediate Measures" in the attachment - New firmware version 1.4-006 or higher will be installed on your battery corpuls cpr (LiPo)