SA-23-08-24-569
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
Philips Azurion, Allura Xper, Integris and MultiDiagnost Eleva
الأجهزة المتأثرة
System Code: 722282, 722281, 722236, 722235, 722234, 722233, 722228, 722227, 722226, 722225, 722224, 722223, 722222, 722221, 722079, 722078, 722068, 722067, 722064, 722063, 722059, 722058, 722044, 722043, 722039, 722038, 722035, 722030, 722029, 722028, 722027, 722026, 722025, 722023, 722021, 722020, 722018, 722017, 722016, 722015, 722013, 722012, 722011, 722010, 722008, 722006, 722005, 722003, 722002, 722001, 708038, 708037, 708036, 708034, 708032 Please refer to the attachment for more details.
الممثل المعتمد
Philips Healthcare Saudi Arabia Ltd
سبب المشكلة
It has been identified that due to the forces applied when moving or turning the monitor of the system, the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may become loose and/or break. If this issue occurs, plastic parts and/or metal bolts of the Cable Hose Carrier(s) may fall and/or part of the cable hose may drop following the detachment of a Cable Hose Carrier.
الإجراء التصحيحي
- Affected systems may continue to be used in accordance with their Instructions for Use (IFU) and the instructions below. - Avoid that operators or bystanders stand underneath the Cable Hose Carriers, especially when moving or turning the system monitor. - Avoid having the patient located underneath the Cable Hose Carriers (e.g., during transport). - If you notice loose or broken bolts and/or broken plastic parts of the Cable Hose Carriers, please contact your local Philips representative immediately. As part of the preventive maintenance cycle, Philips will check the Monitor Ceiling Suspension, including the Cable Hose Carriers, as indicated in the Preventive Maintenance Manual Update attached in Appendix B. Philips is redesigning the Cable Hose Carriers. Once the redesigned cable hose carriers are available, Philips will contact you to schedule a visit to install them in the affected system(s).