تسجيل الدخول
تسجيل جديد
الثلاثاء - رجب 24, 1447
مسجل لدى هيئة الحكومة الرقمية برقم:
20250728714
عن تيقظ
بحث عن منتج
تقديم بلاغ
أسئلة شائعة
تواصل معنا
EN
SA-02-10-25-1078
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
Ceiling Suspension Motorized (CSM) of Philips Diagnostic X-Ray (DXR) systems
الأجهزة المتأثرة
Product Number (REF): 706110, 709031, 712031, 712032, 712034, 712035, 712037 Please refer to the attachment for more details.
الممثل المعتمد
Philips Healthcare Saudi Arabia Ltd
سبب المشكلة
It has been identified by Philips that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues.
الإجراء التصحيحي
Philips has provided the following actions that should be taken by the customer / user in order to prevent risks for patients or users: You may continue to use your system(s) in accordance with the intended use: • Until a Field Service Engineer (FSE) visits your site to inspect the ceiling suspension and implement a hardware correction (if required), Philips recommends using the motorized movement option. • If the force for movement exceeds 110N, the system may experience motor drive errors which restrict the motorized movement. If the motorized movement does not function, please contact your local Philips service representative. A Philips representative will contact you to schedule a time for a Field Service Engineer (FSE) to visit your site to inspect the ceiling suspension and implement a follow-up hardware correction, if required. Manufacturer's Action: Instructions and Correction.
المرفقات
عودة للقائمة