SA-30-09-25-1073
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
CARESCAPE B450, CARESCAPE ONE, B1x5M/B1x5P VSP3.0, B1x5M/B1x5P VSP4.0 Patient Monitors and Portrait Vital Signs Monitor
الأجهزة المتأثرة
Please refer to the attachment.
الممثل المعتمد
General Electric Healthcare Arabia
سبب المشكلة
The potential for loss of monitoring on the affected products when powered by certain batteries.
الإجراء التصحيحي
If possible, use a monitor with a battery that is not within the affected battery population listed in the attachment. However, pending correction by GE HealthCare, you can continue to use a monitor with affected batteries by following the instructions below. 1. Whenever possible, monitors with an affected battery should have mains power connected. 2. If you must use a monitor with an affected battery during patient transport: a. Prior to use, confirm appropriate battery status by checking that: i. No battery failure alarms are indicated on the monitor screen ii. Indicator is green representing sufficient charge level b. Be aware that the monitor may still shut down during patient transport: i. Be prepared to provide alternative methods of assessing the clinical status of the patient until monitor power can be restored ii. Bring a mains power cord during patient transport iii. Because the CARESCAPE ONE requires connection to either a host monitor or to a F0 / F5 FRAME for mains power, bring a spare battery during patient transport Please refer to the specific User Manual for instructions for battery status indicators and battery replacement instructions. Manufacturer's Action: Instructions and Correction.
المرفقات
عودة للقائمة