تسجيل الدخول
تسجيل جديد
الخميس - رجب 26, 1447
مسجل لدى هيئة الحكومة الرقمية برقم:
20250728714
عن تيقظ
بحث عن منتج
تقديم بلاغ
أسئلة شائعة
تواصل معنا
EN
SA-02-09-25-1031
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
Atlan anaesthesia workstations
الأجهزة المتأثرة
Atlan A100, Atlan A100 XL, Atlan A300, Atlan A300 XL, Atlan A350, Atlan A350 XL Please refer to the attachment for serial numbers and more details.
الممثل المعتمد
Draeger Arabia Co Ltd
سبب المشكلة
Possible failure of the automatic ventilation due to impurities (dust) that were introduced for a certain time frame during the manufacturing process of the ventilator motor encoder.
الإجراء التصحيحي
The affected customers will be contacted by Dräger Service to arrange an appointment for replacing the motor. Until then, please continue operating your devices with the customary care by only using the device under permanent supervision in line with the Atlan instructions for use. lf the mechanical ventilation fails, use Man/Spon ventilation mode to ventilate the patient manually as needed. lf necessary, you can downgrade the alarm priority of the "Ventilator error!!!" alarm with "ALARM RESET". Manufacturer's Action: Instructions and Correction.
المرفقات
عودة للقائمة