SA-22-07-25-970
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
Multiple X-ray diagnostic equipments
الأجهزة المتأثرة
Please refer to the attachment.
الممثل المعتمد
Medical Supplies and Services Co Ltd
سبب المشكلة
The contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off.
الإجراء التصحيحي
Please pay attention to the following points until the work by Shimadzu is completed. - For safety reasons, please turn off the breaker of the distribution board to which the power cord of the unit is connected after turning off the power of the unit according to the operation manual. - Do not forget to turn off the breaker of the distribution board until the replacement of the contactor is completed. Shimadzu local authorized representative will take measures to prevent the current from continuing to flow even if the electromagnetic contactor is welded. Manufacturer action: Instructions and correction.