SA-30-05-25-916
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
Philips Azurion Systems R1.x and R2.x
الأجهزة المتأثرة
Model Number (System Product Names): 722063 (Azurion 3 M12) 722064 (Azurion 3 M15) 722067 (Azurion 7 B12/12) 722068 (Azurion 7 B20/15) 722078 (Azurion 7 M12) 722079 (Azurion 7 M20) 722221 (Azurion 3 M12) 722222 (Azurion 3 M15) 722223 (Azurion 7 M12) 722224 (Azurion 7 M20) 722225 (Azurion 7 B12/12) 722226 (Azurion 7 B20/15) 722227 (Azurion 5 M12) 722228 (Azurion 5 M20) 722280 (Azurion 3 M15) 722281 (Azurion 5 M20) 722282 (Azurion 7 M20)
الممثل المعتمد
Philips Healthcare Saudi Arabia Ltd
سبب المشكلة
Software issue potentially resulting in loss of imaging (X-ray) functionality.
الإجراء التصحيحي
• Circulate the attachment to all users of the system so that they are aware of the issue. • Avoid rapidly pressing and releasing the footswitch pedals. • In case X-ray functionality is not available following the situations described in the attchment, perform a cold system restart of the system, as follows: o On the Review Module, press and hold “Power Off”. o Release the button when the indicator light begins to flash. o When the indicator light stops flashing, wait for 10 seconds. o On the Review Module, press and hold “Power On”. NOTE: Do not operate any of the controls while the system is powering on, as this may inhibit the start-up process • Keep the attchment with the documentation of the system until Philips corrects your system. Ensure that the attchment is in a place likely to be seen/viewed. Philips will address the identified issue by implementing a software update in all affected systems. For systems featuring an Interventional Workspot and/or EchoNavigator, Philips will upgrade the Interventional Workspot and/or EchoNavigator software version to ensure compatibility with the updated Philips Azurion system software. Philips representative will contact you to schedule a visit to install the software update once available. Manufacturer Action: Instructions and correction.
المرفقات