SA-20-05-25-900
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
Drive DeVilbiss Healthcare Car Charger Power Cord Accessory
الأجهزة المتأثرة
The car charger cord provided as an accessory with the 125A, 125K series iGo®2 Portable Oxygen Concentrator (POC) system or as a standalone replacement part. Product Name (Catalog Number, UDI): Autoladekabel- Adapter (DV6X-619, 00885304020585) Tragbares Sauerstoffkonzentrator iGo®2 (125A, 00885304022251) (125K, 00885304022244) (125K-XB, 00885304032502) (125K-BT, 00885304033233) (125K-BT-XB, 00885304033240) Please refer to the attachment for the range of impacted lots and serial numbers.
الممثل المعتمد
Bio Standards LLC
سبب المشكلة
During use, the adapter connection may become hot to the touch, and in some cases, cause deformation of the plastic housing.
الإجراء التصحيحي
1. Identify any inventory of the affected car charger in your possession using the identifiers in the attachment. 2. Destroy all affected car charger cords by cutting the cord with wire-cutters or unscrewing the cap and then discarding the car charger parts, refer to the attachment for more detals. Manufacturer action: Removal.
المرفقات