تسجيل الدخول
تسجيل جديد
الثلاثاء - ذو القعدة 22, 1446
عن تيقظ
بحث عن منتج
تقديم بلاغ
أسئلة شائعة
تواصل معنا
EN
SA-01-05-25-878
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
Centricity High Acuity Critical Care (CHA CC) and Centricity High Acuity Anesthesia (CHA A) systems (collectively CHA)
الأجهزة المتأثرة
▪ Centricity High Acuity Critical Care (CHA CC) versions 4.2 through 5.8.3 Patch A. ▪ Centricity High Acuity Anesthesia (CHA A) versions 4.2 through 5.8.3 Patch A.
الممثل المعتمد
General Electric Healthcare Arabia
سبب المشكلة
A potential cybersecurity vulnerability within one of the CHA components.
الإجراء التصحيحي
▪ You can continue to use your CHA applications by ensuring that CHA systems are operated in a secure network environment that is protected from unauthorized intrusion per product documentation. ▪ Customers are encouraged to adopt cybersecurity defense-in-depth industry best practices, implementing necessary technical and operational controls at the perimeter, network, and host levels. ▪ Please ensure that your network administrators and personnel responsible for network security in your facility are made aware of the attachment and the recommended actions. GE HealthCare representative will contact you to arrange for the correction. Manufacturer Action: Instructions and correction.
المرفقات
عودة للقائمة