SA-13-04-25-854
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
Connex Vital Signs Monitor (CVSM), CP150 Electrocardiograph & Connex Spot Monitor Accessory Power Management Stand (APM)
الأجهزة المتأثرة
Please refer to the attachment.
الممثل المعتمد
Baxter AG
سبب المشكلة
The use of an unapproved battery may lead to combustion, fire or device performance issues.
الإجراء التصحيحي
1. Do not use unapproved batteries. Baxter asks customers to only use the approved Baxter / Welch Allyn BATT99 battery, consistent with product labeling due to their safety and performance testing. 2. Inspect all devices. If you have a device identified in the affected product in the attachment, check the device for the correct battery type. See Figure 1 in the attachment for examples of labels from approved batteries labelled with the BATT99 product code, “Welch Allyn”, and the Skaneateles Falls manufacturing facility address. 3. Stop using the device if an unapproved battery is found. 4. Immediately replace the unapproved battery with an approved battery. Manufacturer action: Instructions.