SA-03-04-25-837
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
Philips CT Big Bore
الأجهزة المتأثرة
Product Code (REF): 728242, 728244, 728243 Product Model: Big Bore RT, Brilliance CT Big Bore, Brilliance CT Big Bore Oncology Software: V4.8.0.10430 Please refer to the attachment for more details.
الممثل المعتمد
Philips Healthcare Saudi Arabia Ltd
سبب المشكلة
Motion issues related to Interventional Control that may lead to a collision of the Gantry or table with the operator or patient.
الإجراء التصحيحي
Philips intructs the customers/users to the following: You may continue to use your system(s) in accordance with the intended use and by following the recommendation listed below: • To avoid an unreleased Interventional Control Touch Panel Motion Button, the operator should directly remove their finger from the touch panel motion buttons to release them, rather than dragging or sliding their finger across the panel area. Press any of the Emergency STOP buttons to stop unintended patient table or Gantry motion. • To protect the operator and the patient from the Plan Box not Updating, remain vigilant and confirm table position during procedures. Press any of the Emergency STOP buttons to stop unintended patient table motion. • Refer to Appendix A in the attachment for specific details regarding other issues and recommended actions. Philips will contact the affected customers to schedule a time for a Philips Field Service Engineer (FSE) to visit their sites and install the solution to resolve these issues (reference FCO72800825). Manufacturer action: Instructions and correction.
المرفقات
عودة للقائمة