SA-20-03-25-823
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
Centricity High Acuity Critical Care (CHA CC) and Centricity High Acuity Anesthesia (CHA A) systems (collectively CHA)
الأجهزة المتأثرة
▪ Centricity High Acuity Critical Care (CHA CC) versions 5.8.2 through 5.8.4 with the order module feature enabled. ▪ Centricity High Acuity Anesthesia (CHA A) versions 5.8.2 through 5.8.4 with the order module feature enabled.
الممثل المعتمد
General Electric Healthcare Arabia
سبب المشكلة
A potential issue within the CHA Order module’s Edit Order window. Users may unintentionally reschedule the next administration task to a much later time by inadvertently entering the order ‘End time’ into the 'Next' fields.
الإجراء التصحيحي
You can continue to use your CHA applications in accordance with the User Manuals. Please refer to "Actions to be taken by Customer/User" for more information. A GE HealthCare representative will contact you to arrange for the correction. Manufacturer action: Instructions and correction.
المرفقات