SA-13-03-25-812
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
ABL90 Series - ABL90 FLEX PLUS
الأجهزة المتأثرة
ABL90 FLEX and ABL90 FLEX PLUS analyzers running with RWIN10 versions 1.0 and 1.1.
الممثل المعتمد
Beckman Coulter Saudi Arabia Co Ltd
سبب المشكلة
A potential risk with ABL90 FLEX and ABL90 FLEX PLU analyzers when operating with Radiometer Windows 10 versions (RWIN10 1.0 and 1.1). Specifically, the analyzer software may unexpectedly "freeze" and restart during a sample measurement, causing the analyzer to cease operation during the startup period.
الإجراء التصحيحي
The local Radiometer representative will contact affected customers to discuss upgrading your analyzer to RWIN10 1.2. Please do not use the analyzer without having an alternative solution readily available. Alternative solutions could include using an additional analyzer or sending the sample to a central laboratory. This also involves ensuring that there is enough patient sample for a second measurement, in case the sample is lost due to the analyzer restarting. Manufacturer action: Instructions and correction.