SA-12-02-25-784
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
DxC 500 AU clinical Chemistry Analyzer
الأجهزة المتأثرة
REF | Serial Number | Software Version: C63519 and C63520 | Please refer to "Appendix A" in the attachment | V1.3, V1.4, V1.4.1, V1.4.2, V1.4.3
الممثل المعتمد
Beckman Coulter Saudi Arabia Co Ltd
سبب المشكلة
The affected products cannot recalculate a calculated result when results are rerun within the calculated test.
الإجراء التصحيحي
1) The calculated results can be recalculated manually offline by users based on the formula defined and rerun results from constituent tests. 2) Configure calculated tests within LIS/Middleware, if available. 3) Beckman Coulter recommends sharing the attachment with your laboratory and/or Medical Director. 4) Beckman Coulter will visit you for software update when it is available. Manufacturer action: Instructions and correction.