SA-21-01-25-755
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
Multi Sentry 125 Uninterruptible Power Supply used with with certain Philips CT system models
الأجهزة المتأثرة
Impacted Philips MR Systems Ref, Model (Serial Number(s)): 728143, Incisive CT (500619,500616,500559,500546,510004,500545) 728144, Incisive CT (500519,550024,550184,34114,34115,34113,34138,34172,552066,550341,34439,34455,550509,550598,550327,550245,550339,550342,550340,550032,34173,552037,550228) Please refer to the attachment for the image of Multi Sentry 125 Uninterruptible Power Supply (Riello MST 125kVA UPS).
الممثل المعتمد
Philips Healthcare Saudi Arabia Ltd
سبب المشكلة
There is a potential for the UPS capacitor to fail which may lead to overheating and that it is necessary to replace the capacitor.
الإجراء التصحيحي
It will be scheduled by Philips a time for a Field Service Engineer (FSE) and a Riello representative to replace the affected products. Manufacturer action: correction.
المرفقات