SA-20-01-25-750
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
Medima Infusion Pump
الأجهزة المتأثرة
Medima Infusion Pump
الممثل المعتمد
ALKAMAL IMPORT OFFICE
سبب المشكلة
A software issue was detected during internal testing by the manufacturer. The issue only impacts infusions that meet both conditions below: - The infusion is programmed using the “Profile” infusion mode with units of mL/min or mL/24h - The user titrates the dose rate after the infusion is started
الإجراء التصحيحي
If you use software version 3.1.87 and older, do not use dosing units of mL/min or mL/24h with the “Profile” infusion mode and refrain from making dose rate changes when this mode is selected. In “Profile” infusion mode, please use only ml/h units until you update your software version to 3.1.88 or above, follow "Required Actions for Users" in the attachment". Medima has developed a new software version (3.1.88 and above) that resolves the issue. Manufacturer action: instructions and correction.
المرفقات