SA-04-11-24-657
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
Interventional Angiography System
الأجهزة المتأثرة
X-ray interventional systems: INFX8000C; INFX-8000V Combined with XIDF-SFL801: SPOT Fluoro kit (option). Applicable shipped date: shipped from Sep. 2023 to Apr. 2024.
الممثل المعتمد
Gulf Medical Co.
سبب المشكلة
When SPOT Fluoro is first turned on, the default ROI is selected, and the collimator blades move to the boundary of the ROI, and the static image is overlayed around the ROI, covering the collimated area. If customer changes the ROI, the collimator blades and static images are adjusted accordingly. However, when the acquisition program is changed, the collimator returns to the default ROI position which may be larger than the specified one, leaving the X-ray exposed area partially covered by the static image. This is a software problem with the SPOT Fluoro program and will not occur if the function is not used.
الإجراء التصحيحي
A software update is being developed by Canon to correct the problem. Customers will be contacted when it is available. Until the software update is available customers are advised to not use the Spot Fluoro function. Should this problem occur before the modified software is installed on your system, please perform the mentioned method in the attachment to avoid the event occurrence. Manufacturer Action: Instructions and Correction.