SA-17-12-24-708
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
Certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems
الأجهزة المتأثرة
Please refer to "Affected Product Details" in the attachment.
الممثل المعتمد
General Electric Healthcare Arabia
سبب المشكلة
The affected products could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms.
الإجراء التصحيحي
Please follow actions to be taken by Customer/User in the attachment. A GE HealthCare Field Engineer will inspect the potentially affected systems in Appendix A of the attachment to determine if the detector support mechanisms are intact and functioning correctly. GE HealthCare will apply warning labels to the potentially affected systems and will provide you with updated instructions on how to correctly relocate your system, so the detector mounting mechanism is not compromised. Manufacturer Action: Instructions and Correction.
المرفقات