SA-26-11-24-688
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
Corpuls AED
الأجهزة المتأثرة
Battery corpuls AED Refer to the attachment for more details.
الممثل المعتمد
Al Khateeb United Trading And Contracting Company
سبب المشكلة
The affected devices may perform an unplanned restart with self-test during operation.
الإجراء التصحيحي
- Affected batteries can be identified by performing the pressure point test - see the attachment (Identifying Affected Devices). - If you have identified an affected product by means of the test procedure described in the attachment, decommission the affected battery and contact your authorized sales- and service partner. A permanent improvement of conspicuous devices can be achieved by exchanging the battery or by modifying the corpuls aed device in the area of the battery interface board. Both solutions can be implemented independently to correct the faulty behavior of the device, one of the two solutions will be prepared for you by your authorized sales- and service partner. Manufacturer Action: Instructions and Correction.