SA-04-11-24-659
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
ARGUS 717V and ARGUS 718V
الأجهزة المتأثرة
Device Model/Catalogue/part number(s): ARGUS 717V (REF 601268) ARGUS 718V (REF 601349)
الممثل المعتمد
ALJEEL MEDICAL COMPANY
سبب المشكلة
The problem involves the power board/mainboard female-on-board Micro-Match connector, the flat ribbon cable, and the Micro-Match male connector. If any of these components are defective, the connection between the mainboard and the power board can be compromised.
الإجراء التصحيحي
Service technicians / service providers information: • The flat ribbon cable (REF: 601342 - flat ribbon cable 24-pole) shall always be removed from the female connector by holding and pulling the male connector, not by tugging on the wires. • If the cable of the identified device has been subjected to high mechanical stress at any point during the pump's lifetime, either through stretching during repairs or improper disconnection, the cable should be replaced with a new one (REF: 601342 - flat ribbon cable 24-pole). • In the case of any future service, it is recommended to replace the flat ribbon cable with a new one (REF: 601342 - flat ribbon cable 24-pole), once it has been disconnected from the power board/mainboard. Manufacturer Action: Correction.