SA-03-11-24-654
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
SIGNA Architect, SIGNA Architect AIR, Discovery MR750w 3.0T, and SIGNA PET/MR systems
الأجهزة المتأثرة
SIGNA Architect, SIGNA Architect AIR, Discovery MR750w 3.0T, and SIGNA PET/MR systems. Please refer to the attachment for more details.
الممثل المعتمد
General Electric Healthcare Arabia
سبب المشكلة
The gradient coils for the affected products, under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels may exceed 99dB.
الإجراء التصحيحي
You can continue to use your MR system by following the instructions below: Ensure hearing protection with a Noise Reduction Rating (NRR) of at least 33dB is used during exams on these systems. Please post the attachment in your facility on or near the MR operator console.