SA-03-11-24-655
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
Proteus XR/a
الأجهزة المتأثرة
Proteus XR/a Systems (GTIN: 00840682120777). Please refer to the attachment for affected system IDs.
الممثل المعتمد
General Electric Healthcare Arabia
سبب المشكلة
The steel cables in the Wall Stand for certain Proteus XR/a systems serviced by GE HealthCare have not been replaced at the intervals detailed in the Preventative Maintenance Action List (GEHC Service Manual 2273022-100).
الإجراء التصحيحي
You can continue to use the system for table exams without limitations. • You can continue to use the wall stand for exams where the patient does not have any part of their body underneath the bucky. • Do not perform exams where the patient has any body part underneath the bucky until the Wall Stand steel cable has been inspected and if necessary, has been replaced. A GE HealthCare representative will contact you to arrange for the correction.