SA-28-10-24-647
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
Universal remote device
الأجهزة المتأثرة
Item number: 100925A0
الممثل المعتمد
Gulf Medical CO LTD
سبب المشكلة
Under certain conditions, it has been identified that an issue might prevent the device from performing as intended. It has been determined that with the above Getinge IR remote control in some cases no function occurs and therefore no movement of the operating table is possible.
الإجراء التصحيحي
In the event no function of Getinge IR Remote Control 100925A0 occurs the user has the possibility to use the override control panel of the operating table, refer to the attachment for more details. The user has also the possibility to use a corded control device. A Software update that will correct this issue has been developed. You will be contacted by your Getinge sales or service representative to plan for the update of your device.