MERIVAARA Q-FLOW SURGICAL LIGHT
الأجهزة المتأثرة
Q-FLOW SOLO, Q-FLOW DUO, Q-FLOW TRIO, Q-FLOW QUAD, Q-FLOW MOBILE The above listed Merivaara Q-Flow lights manufactured before June 2023.
الممثل المعتمد
Omar Saeed Alamoudi Sons Co LLC
سبب المشكلة
Due to an assembly error, the device’s support yokes can become loose from their fixings. In the worst case, if these fixing screws become completely loose, the luminaire may fall off its fixture.
الإجراء التصحيحي
Merivaara recommends that the user follows the instructions for use and inspects the luminaire daily when cleaning and pays attention to the joints of the support yokes shown in Figure 2 in the attachment. If looseness is found at these joints during the daily inspection, the luminaire should be taken out of use and repaired by a qualified service technician according to the repair instructions. The corrective actions include the repair of the fixing of the luminaire’s support yoke/yokes . The repair includes the addition of screw lock to the four screws used in the support yoke assembly. Depending on the model of the luminaire, there are 1–3 support yokes to be repaired under one product name. The repair shall be carried out according to the instructions in the attachment.