SA-08-10-24-618
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
Atlan Anaesthesia Workstations
الأجهزة المتأثرة
Affected Devices: Atlan A100, Atlan A300, Atlan A300 XL, Atlan A350, Atlan A350 XL Multiple Model Numbers and serial numbers Please refer to the attachment.
الممثل المعتمد
Draeger Arabia Co Ltd
سبب المشكلة
Possible failure of the automatic ventilation due to an error in the motor electronics.
الإجراء التصحيحي
The affected customers will be contacted by Dräger Service to arrange an appointment for replacing the motor. Until then, please continue operating your devices with the customary care by only using the device under permanent supervision in line with the Atlan instructions for use. lf the mechanical ventilation fails, use Man/Spon ventilation mode to ventilate the patient manually as needed. lf necessary, you can downgrade the alarm priority of the "Ventilator error!!!" alarm with "ALARM RESET".