SA-30-09-24-608
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
Zenition 50 and Zenition 70 systems
الأجهزة المتأثرة
Product Description - Product Code - Product UDI Zenition 50 - 718096 - (01)00884838091535(21) Zenition 70 - 718133 - (01)00884838091528(21)
الممثل المعتمد
Philips Healthcare Saudi Arabia Ltd
سبب المشكلة
The 5-Ampere Mains Control Unit (MCU) board fuse may blow out during the system start-up or during an ongoing procedure. When this issue occurs, the system will display a combination of error messages M326, M050 and M501 during a system start-up and M370, and M501 during an ongoing procedure.
الإجراء التصحيحي
Affected systems may continue to be used in accordance with their intended use and Instructions for Use (IFU). The 5-Ampere MCU board fuse will be replaced by Philips with a 10-Ampere MCU board fuse in all affected systems (reference FCO71800109).