SA-15-09-24-593
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
Giraffe OmniBed and Giraffe OmniBed Carestation devices
الأجهزة المتأثرة
All Giraffe OmniBed Carestations manufactured between January 2018 and November 2021 and All Giraffe OmniBeds and Giraffe OmniBed Carestations that have had heater doors replaced.
الممثل المعتمد
General Electric Healthcare Arabia
سبب المشكلة
The screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose. If the doors become loose, the system activates a high priority alarm, and the canopy movement will stop.
الإجراء التصحيحي
You may continue to use your devices. Please always pay attention to product alarms and follow the instructions in the product user manual. Download the Service Manual Addendum (5971733) from the GE HealthCare Customer Documentation Portal (refer to the attachment). Follow the instructions in the downloaded service manual addendum. If you identify a damaged heater door please indicate this on the response form in the attachment and GE HealthCare will provide a replacement at no cost to you.