SA-13-09-24-588
التفاصيل
الرقم المرجعي
الشركة الصانعة
اسم الجهاز الطبي
Enterprise Imaging XERO Viewer
الأجهزة المتأثرة
Enterprise Imaging XERO Viewer (UDI-DI: 05400874000710) All versions
الممثل المعتمد
Gulf Medical Co.
سبب المشكلة
An issue related to orientation of 3D volumes using reference lines, active target localization, and series linking across multiple planes. A coding error has been identified in the initialization of pixel spacing in the XERO Viewer.
الإجراء التصحيحي
When making any diagnosis and/or treatment decision regarding findings associated with the location of the reference line or active target in XERO Viewer, it is suggested to follow "Recommended actions to be taken" in the attachment. AGFA HealthCare will provide the software corrections for the affected versions. A patch is available for each of the following XERO Viewer version streams (8.4.x, 8.3.x, 8.2.x, 8.1.4.x & 8.1.2.x), please refer to attachment for more details.